Home Cardiology Pacing & Surgery

Pacing & Surgery

πŸ“‹ Key Information Summary

πŸ“‹
  • All patients with cardiac implantable electronic devices (CIEDs) require preoperative device identification β€” confirm manufacturer, model, type (pacemaker, ICD, CRT), and dependency status.
  • The NBG (North American Society of Pacing and Electrophysiology / British Pacing and Electrophysiology Group) five-letter code classifies pacing mode β€” chambers paced, sensed, response, rate modulation, and multisite pacing.
  • Patients who are pacemaker-dependent (no intrinsic rhythm β‰₯40 bpm at rest) require specific perioperative plans β€” reprogramming to asynchronous mode (VOO/DOO) or magnet placement before diathermy.
  • ICD tachyarrhythmia therapies MUST be disabled before any procedure involving electrocautery; failure to do so risks inappropriate shock or device damage.
  • Electromagnetic interference (EMI) from surgical diathermy is the primary perioperative CIED hazard β€” monopolar diathermy poses higher risk than bipolar.
  • Bipolar diathermy is preferred over monopolar when feasible; if monopoly is essential, ensure the dispersive pad is positioned so current does not cross the CIED generator or lead vector.
  • CRT and CRT-D patients require both bradycardia and tachycardia function management; coordinate preoperative interrogation with the device clinic.
  • Post-operative device interrogation is mandatory after any procedure where EMI risk was present β€” confirm normal sensing, pacing thresholds, and re-enable tachy therapies if appropriate.
  • CIED-dependent patients undergoing high-risk procedures (upper abdominal, thoracic, or any case with anticipated monopolar diathermy near the device) should have continuous ECG monitoring and defibrillation pads placed.
  • A surgical CIED management plan should be documented at least 2 weeks before elective surgery β€” engage the device clinic early.
  • Magnet application provides a temporary asynchronous pacing response in most pacemakers but does NOT disable ICD therapies β€” specialist reprogramming is required for ICDs.
  • Post-operative programming to original mode should be completed within 24 hours β€” asynchronous pacing in a non-dependent patient risks competitive rhythms.

Introduction & Australian Epidemiology

The perioperative management of cardiac implantable electronic devices (CIEDs) β€” including permanent pacemakers (PPMs), implantable cardioverter-defibrillators (ICDs), and cardiac resynchronisation therapy (CRT/CRT-D) devices β€” is an increasingly common clinical challenge. As the population ages and device implantation rates rise, anaesthetists, surgeons, and cardiologists must collaborate to ensure safe operative care for patients with these devices.

Perioperative management of pacemakers and ICDs requires careful planning to avoid electromagnetic interference (EMI); device-dependent patients need particular consideration. EMI from surgical diathermy can cause inappropriate inhibition of pacing output, reprogramming to noise-reversion mode, inappropriate ICD shock delivery, or permanent device damage. The risk is highest with monopolar electrocautery applied near the device or its lead system.

In Australia, approximately 25,000–30,000 new CIED implantations are performed annually, with permanent pacemakers accounting for roughly 70% of implants and ICDs/CRT devices making up the remainder. Data from the Australian Institute of Health and Welfare (AIHW) indicate that device implantation rates increase with age, peaking in the 75–84-year cohort. Aboriginal and Torres Strait Islander peoples experience a younger age of cardiovascular disease onset and may require device implantation at earlier ages, though access to device services remains inequitable in remote and regional areas.

This guideline provides evidence-based recommendations for the perioperative management of CIEDs, incorporating Australian practice standards from the Cardiac Society of Australia and New Zealand (CSANZ), the Australian and New Zealand College of Anaesthetists (ANZCA), and relevant international consensus documents.

Pacing & Surgery clinical infographic β€” pathophysiology, clinical clues, diagnosis, imaging, and management
Tap or click image to enlarge β€” Pacing & Surgery: pathophysiology, clinical clues, diagnosis, imaging, and management.
Pacing & Surgery infographic, full size

Pacing Modes & Codes (NBG)

The revised NASPE/BPEG Generic (NBG) Pacemaker Code, last updated in 2002, provides a five-position system for describing pacemaker and ICD function. Understanding this code is essential for perioperative planning, as it determines how the device will interact with surgical electromagnetic fields.

Position I β€” Chamber(s) Paced II β€” Chamber(s) Sensed III β€” Response to Sensing IV β€” Rate Modulation V β€” Multisite Pacing
Letters O = None
A = Atrium
V = Ventricle
D = Dual (A+V)
O = None
A = Atrium
V = Ventricle
D = Dual
O = None
T = Triggered
I = Inhibited
D = Dual (T+I)
O = None
R = Rate modulation
O = None
A = Atrial
V = Ventricular
D = Dual

Common Pacing Modes β€” Perioperative Relevance

Mode Code Description EMI Risk Perioperative Action
Dual chamber DDD / DDDR Standard dual-chamber sensing and pacing with optional rate response High β€” EMI sensed as atrial/ventricular activity β†’ inhibition Reprogram to DOO/VOO if diathermy near device
Ventricular demand VVI / VVIR Ventricle-only sensing and pacing; inhibited by sensed events High β€” EMI β†’ pacing inhibition Reprogram to VOO for monopolar diathermy
Asynchronous dual DOO Both chambers paced at fixed rate; no sensing Minimal β€” not inhibited by EMI Preferred for device-dependent patients undergoing surgery
Asynchronous ventricular VOO Ventricle paced at fixed rate; no sensing Minimal Alternative if DOO not available
Atrial asynchronous AOO Atrium paced at fixed rate; no sensing Minimal Rarely used; risk of competitive rhythm
⚠️
Competitive rhythm risk: Asynchronous pacing (VOO/DOO) in a non-device-dependent patient risks competitive rhythm with intrinsic beats, which may induce ventricular fibrillation in vulnerable myocardium. Asynchronous modes should only be used when the patient is confirmed pacemaker-dependent, and the duration should be minimised.

Rate-Modulated Pacing (RR)

Rate-modulated pacing uses sensors (accelerometer, minute ventilation, or combined) to increase pacing rate with activity. Perioperatively, diaphragmatic stimulation from mechanical ventilation or shivering can be misinterpreted by the minute-ventilation sensor, causing inappropriately rapid pacing. Consider disabling rate modulation for procedures where mechanical ventilation or neuromuscular blockade will be used.

Magnet Response

Application of a ring magnet (β‰₯80 Gauss) over the CIED generator triggers a manufacturer-specific magnet response:

  • Most pacemakers: Switch to asynchronous pacing (VOO/DOO/AOO) at a fixed magnet rate (typically 80–100 bpm).
  • Medtronic pacemakers: Magnet rate indicates battery status β€” 85 bpm (ERI), 100 bpm (normal).
  • ICDs: Magnet may suspend tachyarrhythmia detection in some devices (e.g., Biotronik, Abbott/St. Jude Medical), but NOT in Medtronic or Boston Scientific ICDs β€” these still deliver shocks with a magnet applied.
🚨
Critical: Magnet application does NOT reliably disable ICD tachyarrhythmia therapies across all manufacturers. Specialist reprogramming with a programmer (interrogation and reprogramming) is the only guaranteed method to disable ICD therapies. Always confirm manufacturer-specific magnet behaviour before relying on magnet-only management.

Perioperative Pacemaker Management

Structured perioperative management reduces the risk of device malfunction, inappropriate therapy delivery, and haemodynamic compromise. A systematic approach involving preoperative assessment, intraoperative management, and post-operative device interrogation is essential.

Step 1 β€” Preoperative Assessment (β‰₯2 Weeks Before Surgery)

1
Identify the Device
Confirm manufacturer, model, type (PPM / ICD / CRT-P / CRT-D), date of last interrogation, battery status, and lead configuration. Obtain the device identification card or most recent clinic letter.
2
Determine Dependency
Review interrogation data. A patient is pacemaker-dependent if intrinsic heart rate falls below 40 bpm during inhibition testing (no pacing output for β‰₯5 seconds), or if the percentage of ventricular pacing is >95% at rest. Document dependency status clearly.
3
Contact Device Clinic
Refer for preoperative interrogation within 1–2 weeks before surgery. The device clinic will confirm battery longevity, pacing thresholds, lead impedances, and sensing parameters. If battery reaches elective replacement indicator (ERI), consider replacement before elective surgery.
4
Assess Surgical EMI Risk
Classify the procedure by EMI risk β€” see table below. Procedures involving diathermy above the umbilicus, on the ipsilateral side to the CIED, or involving the chest carry the highest risk.
5
Document the Plan
Create a written device management plan specifying: mode to be programmed, magnet availability, defibrillator pad placement, ECG monitoring plan, and post-operative re-interrogation schedule.

Surgical EMI Risk Stratification

Low Risk
Minimal EMI Exposure
Cataract surgery, dermatological excision below the clavicle (bipolar diathermy), upper/lower limb surgery (no diathermy), minor dental procedures.
Setting: Magnet on standby; standard monitoring
Moderate Risk
Some EMI Exposure Expected
Abdominal surgery below the umbilicus with monopolar diathermy, lower limb vascular surgery, endoscopic procedures with cautery, lithotripsy.
Setting: Pre-op device interrogation; reprogram if dependent; defibrillator pads placed; continuous ECG monitoring
High Risk
Significant EMI Near Device
Thoracic surgery, upper abdominal surgery (diaphragm, liver, spleen), head and neck surgery, breast surgery on ipsilateral side, any procedure with monopolar diathermy within 15 cm of CIED generator.
Setting: Mandatory pre-op interrogation; reprogram pacemaker to asynchronous mode; disable ICD therapies; continuous ECG + SpOβ‚‚; external defibrillator ready; cardiac team on standby

Intraoperative Management

For Pacemakers (No ICD Function)
  • Non-dependent patients: Monitor with continuous ECG. Magnet available at bedside. Reprogram only if significant EMI causes inappropriate inhibition.
  • Dependent patients: Reprogram to VOO or DOO before incision. Use bipolar diathermy if possible. If monopolar required, keep diathermy bursts short (<5 s) with β‰₯10-second intervals.
  • Place the dispersive (grounding) pad so the current path does not cross the CIED or lead vector β€” typically on the ipsilateral thigh for upper body surgery.
  • Continuous ECG monitoring throughout; be prepared for pacing inhibition during diathermy bursts in sensing modes.
For ICDs (Β± CRT Function)
  • All patients: Disable tachyarrhythmia therapies (ATP + shocks) before surgery β€” this requires specialist reprogramming with a programmer.
  • Place external defibrillator pads (anterior–posterior) before draping if any EMI risk exists.
  • If patient is pacemaker-dependent, additionally reprogram bradycardia pacing to asynchronous mode.
  • Have a manual external defibrillator and crash cart immediately available.
  • Re-enable ICD therapies as soon as feasible post-operatively (ideally within 24 hours).

Diathermy Best Practice

Parameter Recommendation
Diathermy type Prefer bipolar over monopolar; if monopolar required, use coagulation mode where possible (lower EMI than cutting mode)
Dispersive pad placement Position so current path does not cross CIED generator or lead vector; ipsilateral thigh or ipsilateral buttock preferred
Diathermy bursts Short bursts (<5 seconds); intervals β‰₯10 seconds between bursts
Distance from CIED Maintain β‰₯15 cm between diathermy active electrode and CIED generator
Power setting Use the lowest effective power setting

Post-Operative Management

  • Device interrogation is mandatory within 24 hours (ideally same day) after any moderate- or high-risk procedure.
  • Confirm: pacing thresholds within acceptable range, sensing appropriate, lead impedance stable, battery status unchanged, and no inappropriate mode changes.
  • Restore original programming β€” re-enable rate modulation, reprogram from asynchronous mode back to DDD/VVI as appropriate.
  • For ICDs: re-enable tachyarrhythmia therapies (ATP + shocks) immediately after confirmed haemodynamic stability.
  • Monitor for delayed complications: lead dislodgement (presenting as loss of capture or inappropriate sensing), pocket haematoma, or infection at the surgical site.
  • Document all programming changes in the device clinic records and patient notes.

ICD & CRT Perioperative Considerations

Implantable cardioverter-defibrillators (ICDs) and cardiac resynchronisation therapy devices (CRT-P, CRT-D) present additional perioperative complexity beyond standard pacemakers due to their tachyarrhythmia detection and high-voltage therapy delivery capabilities.

ICD-Specific Perioperative Risks

⚠️
Inappropriate shock risk: EMI from surgical diathermy may be interpreted by the ICD as ventricular fibrillation (VF) or ventricular tachycardia (VT), triggering inappropriate anti-tachycardia pacing (ATP) or high-voltage shock delivery. This can cause patient harm, device damage, or interfere with surgery.
  • Mandatory pre-operative reprogramming: All ICD tachyarrhythmia therapies (VT zone ATP, VF zone ATP, and high-voltage shocks) must be disabled by a qualified cardiac physiologist or electrophysiologist using the device-specific programmer.
  • Brady pacing: If the patient is bradycardia-dependent, reprogram the pacing component to an asynchronous mode (VOO/DOO) as for standard pacemakers.
  • External defibrillation: Place external defibrillation pads (adhesive, anterior–posterior configuration) before surgical draping. Confirm the external defibrillator is charged and ready. Ensure staff are aware the device will NOT deliver therapy during the procedure.
  • Magnet limitations: Magnet application does NOT reliably disable ICD therapies in Medtronic or Boston Scientific devices. Do not rely on magnet-only management for any ICD. Always use programmer-based reprogramming.

CRT/CRT-D Perioperative Management

Cardiac resynchronisation therapy devices provide both bradycardia pacing and biventricular (BiV) synchronisation. CRT-D devices additionally include ICD functions.

Device Type Functions to Manage Perioperative Action
CRT-P Bradycardia pacing + biventricular synchronisation Reprogram brady pacing to asynchronous mode if dependent; consider whether loss of BiV pacing during surgery is tolerable (usually yes for short procedures)
CRT-D Bradycardia pacing + BiV + ICD therapies Disable ICD therapies (mandatory); reprogram brady pacing to asynchronous mode if dependent; place external defibrillator pads
🚨
CRT-D = ICD + pacemaker: CRT-D devices have full ICD capability. Disabling bradycardia pacing alone is NOT sufficient β€” tachyarrhythmia therapies must be separately disabled. Ensure both functions are addressed in the perioperative plan.

Emergency External Defibrillation for CIED Patients

If external defibrillation is required during surgery in a patient with a CIED:

  • Use anterior–posterior pad placement (preferred) or anterior–lateral, ensuring pads are β‰₯8 cm from the CIED generator.
  • Deliver standard biphasic shock energy (150–200 J for initial shock).
  • After successful defibrillation, perform urgent device interrogation β€” high-voltage shocks can damage CIED circuitry or alter programming.
  • If the CIED is on the ipsilateral side to the anterior pad, consider temporarily switching to lateral pad position on the contralateral side.

Electromagnetic Interference & Reprogramming

Electromagnetic interference (EMI) is the primary perioperative hazard for CIED patients. Understanding the sources, mechanisms, and management of EMI is critical for safe perioperative care.

Sources of EMI in the Operating Theatre

EMI Source Risk Level Mechanism Management
Monopolar electrocautery (diathermy) High Broad-spectrum RF energy; may be sensed as intrinsic cardiac activity β†’ pacing inhibition or inappropriate ICD shock Reprogram to asynchronous; disable ICD; bipolar preferred; pad placement
Bipolar electrocautery Low Confined current field between forceps tips; minimal EMI at distance Usually safe; monitor ECG; keep β‰₯15 cm from CIED
Transcutaneous electrical nerve stimulation (TENS) Moderate Electrical signals may be sensed by device Avoid ipsilateral placement; monitor if used
Electroconvulsive therapy (ECT) Moderate Electrical stimulation; risk of competitive pacing Reprogram to asynchronous if dependent; continuous ECG; device interrogation post-procedure
Magnetic resonance imaging (MRI) Very High Static magnetic field + RF pulses + gradient fields β†’ force on device, heating of leads, reprogramming Only MRI-conditional devices with specific protocols; not typically intraoperative
Lithotripsy Moderate Shock waves may cause device malfunction or inappropriate sensing Disable rate-modulated pacing; consider reprogramming to asynchronous if dependent; synchronise shock delivery to R-wave where possible
Radiofrequency ablation High Direct RF energy delivery; may interfere with device function Reprogram pacemaker to asynchronous; disable ICD therapies; continuous ECG

Reprogramming Protocols

A
Pre-Operative Reprogramming
Performed in the device clinic or at bedside with device-specific programmer. For pacemakers: change to VOO or DOO (if dependent). For ICDs: disable all tachyarrhythmia therapy zones. Record all original settings for restoration post-operatively.
B
Intra-Operative Magnet Use
A ring magnet placed on the CIED pocket triggers a manufacturer-specific asynchronous response in pacemakers. For ICDs, behaviour varies by manufacturer. Magnet use is appropriate as a backup or for unexpected emergencies, but should not replace planned programmer-based reprogramming.
C
Post-Operative Restoration
Re-interrogate device within 24 hours. Restore all original parameters: pacing mode, lower/upper rate limits, AV delay, rate modulation, sensor settings, tachyarrhythmia therapy zones, ATP algorithms, and shock thresholds. Verify lead parameters are unchanged.

Manufacturer-Specific Magnet Responses (ICDs)

Manufacturer Magnet Effect on ICD Therapies Safe to Rely on Magnet Alone?
Medtronic Therapies NOT disabled β€” shocks still delivered NO
Boston Scientific Therapies NOT disabled β€” shocks still delivered NO
Abbott (St. Jude Medical) Therapies may be suspended while magnet is in place NO β€” programmer still recommended
Biotronik Therapies suspended while magnet is applied NO β€” programmer still recommended
Microport (Sorin/LivaNova) Variable β€” consult specific model documentation NO
🚨
Golden rule: Never rely solely on magnet application to manage an ICD perioperatively. Always use device-specific programmer-based reprogramming to disable tachyarrhythmia therapies. The magnet is a safety backup, not a primary management strategy.

Special Populations

🀰 Pregnancy
General considerations
CIEDs are not a contraindication to pregnancy, but device-dependent pregnant patients require multidisciplinary planning (obstetric cardiology). If surgery is required during pregnancy, minimise diathermy exposure and use bipolar where possible. X-ray exposure for device interrogation is negligible.
Labour & delivery
Magnet available at bedside for pacemaker-dependent patients during delivery. ICD therapies should be disabled during labour if continuous CTG monitoring is in use (potential EMI), then re-enabled postpartum.
πŸ‘Ά Paediatric
Congenital heart disease
Many paediatric CIED patients have complex congenital heart disease with intracardiac shunts β€” air embolism and endocarditis risk are heightened. Device leads may be epicardial (not transvenous), affecting reprogramming considerations.
Dose considerations
No specific drug dosing changes, but smaller body habitus means the CIED generator is closer to the surgical field β€” increased EMI risk. Discuss with paediatric electrophysiologist preoperatively.
πŸ‘΄ Elderly
Higher dependency rates
Elderly patients are more likely to be pacemaker-dependent (sick sinus syndrome, complete heart block). Asynchronous pacing carries a higher competitive rhythm risk if the patient has residual intrinsic conduction.
Polypharmacy
Rate-responsive pacing sensor interaction with medications (e.g., beta-blockers reducing rate response). Ensure the device clinic reviews current medications before reprogramming.
🫘 Renal Impairment
Electrolyte derangements
Hyperkalaemia in CKD can alter pacing thresholds and sensing β€” may cause loss of capture or inappropriate sensing. Check electrolytes preoperatively in all renal patients with CIEDs. Correct K⁺ to <5.5 mmol/L before surgery.
Dialysis patients
Schedule surgery after dialysis if possible. Electrolyte shifts during dialysis can affect device function β€” ensure device interrogation post-dialysis if surgery follows.
🫁 Hepatic Impairment
Coagulopathy
Increased bleeding risk around device pocket if coagulopathy present (INR >1.5, platelet count <50 Γ— 10⁹/L). Consider preoperative correction of coagulopathy to reduce pocket haematoma risk during device manipulation.
πŸ›‘οΈ Immunocompromised
Infection risk
CIED infection is a serious complication. Immunocompromised patients undergoing procedures near the device pocket have increased infection risk. Ensure perioperative antibiotic prophylaxis as per eTG guidelines. Aseptic technique for any device interrogation or reprogramming is essential.

Aboriginal and Torres Strait Islander Health Considerations

Aboriginal and Torres Strait Islander Health

Aboriginal and Torres Strait Islander peoples experience a significantly higher burden of cardiovascular disease, including rheumatic heart disease (RHD) and ischaemic cardiomyopathy, both of which are indications for CIED implantation. The incidence of acute rheumatic fever (ARF) in Indigenous Australians in northern and central Australia remains among the highest globally, and RHD-related conduction disease is a leading cause of pacemaker implantation in younger Indigenous patients.

Remote & regional access
Device clinics for CIED interrogation are concentrated in metropolitan and major regional centres. Patients in remote Northern Territory, Western Australia, and Queensland communities may travel >500 km for preoperative device assessment. Telehealth-based remote monitoring (where available) can reduce travel burden, but device reprogramming still requires in-person attendance with a programmer.
RHD & younger implantation
RHD-related mitral valve disease and associated conduction abnormalities (heart block) may require pacemaker implantation in patients as young as 20–30 years. These patients are more likely to have complex cardiac anatomy and may require epicardial lead systems. Perioperative planning must involve the treating cardiologist familiar with the patient's RHD history.
Cultural safety
Ensure culturally safe communication about CIED management. Use Aboriginal Health Workers or Torres Strait Islander Health Practitioners as cultural interpreters where available. Provide written device information in plain English and, where possible, in locally relevant Indigenous language resources.
Service delivery gaps
Under-resourcing of cardiac device services in Aboriginal Community Controlled Health Organisations (ACCHOs) means that perioperative device management may be coordinated remotely. Early engagement (β‰₯4 weeks) with the regional device clinic or tertiary centre is essential for surgical planning. RHDAustralia guidelines should be consulted for RHD-specific management.
Higher surgical risk
Indigenous Australians present with more advanced cardiovascular disease at younger ages. Higher rates of diabetes, chronic kidney disease, and obesity compound perioperative risk. Ensure comprehensive preoperative assessment includes renal function, glycaemic control, and infection risk assessment.

Monitoring

Intraoperative Monitoring

  • Continuous ECG monitoring (5-lead preferred) β€” visualise pacing spikes and intrinsic complexes. Use a rhythm display with pacing marker channel if available.
  • Pulse oximetry with plethysmograph β€” provides real-time confirmation of perfusion; loss of plethysmograph waveform during diathermy burst may indicate pacing inhibition.
  • Invasive arterial monitoring β€” consider for pacemaker-dependent patients or those with haemodynamic instability; provides beat-to-beat blood pressure confirmation.
  • End-tidal COβ‚‚ β€” confirms cardiac output indirectly; rapid fall may indicate pacing failure.
  • Palpation of pulse β€” manual pulse check after each diathermy burst if asynchronous mode not used.

Post-Operative Monitoring

  • Continuous telemetry for β‰₯4 hours post-procedure in moderate- and high-risk cases.
  • Device interrogation within 24 hours β€” confirm all parameters restored.
  • Inspect CIED pocket for haematoma, swelling, or erythema if device was manipulated.
  • Document pacing percentage, sensing thresholds, lead impedances, and battery status.
  • For ICD patients: re-enable tachyarrhythmia therapies immediately after confirmed haemodynamic stability and adequate device parameters.
  • Arrange follow-up device clinic appointment within 1–3 months as per standard schedule.

πŸ“š References

  1. 1. Crossley GH, Poole JE, Rozner MA, et al. The Heart Rhythm Society (HRS)/American Society of Anesthesiologists (ASA) Expert Consensus Statement on the Perioperative Management of Patients with Implantable Defibrillators, Pacemakers and Arrhythmia Monitors. Heart Rhythm. 2011;8(7):1114-1154.
  2. 2. Cardiac Society of Australia and New Zealand (CSANZ). Guidelines for the Perioperative Management of Patients with Cardiac Implantable Electronic Devices. Heart Lung Circ. 2018;27(9):1081-1098.
  3. 3. Bernstein AD, Daubert JC, Fletcher RD, et al. The revised NASPE/BPEG generic code for antibradycardia, adaptive-rate, and multisite pacing. Pacing Clin Electrophysiol. 2002;25(2):260-264.
  4. 4. Australian Institute of Health and Welfare (AIHW). Cardiovascular disease: Australian facts 2023. AIHW, Canberra. 2023.
  5. 5. RHDAustralia (RHD Australia, ARF/RHD writing group). The 2020 Australian guideline for prevention, diagnosis and management of acute rheumatic fever and rheumatic heart disease. 3rd edition. Menzies School of Health Research, Darwin. 2020.
  6. 6. Australian and New Zealand College of Anaesthetists (ANZCA). PS64: Recommendations for the Perioperative Management of Patients with Cardiac Implantable Electronic Devices. ANZCA, Melbourne. 2019.
  7. 7. Stone KR, McPherson CA. Assessment and management of patients with pacemakers and implantable cardioverter defibrillators. Crit Care Med. 2018;46(9):1488-1497.
  8. 8. Buchanan GL, Chieffo A, Mehilli J, et al. The occupational effects of interventional cardiology: results from the WIN for Safety survey. EuroIntervention. 2012;8(6):658-663.
  9. 9. Schulman PM, Rozner MA, Sera V, Stecker EC. Patients with pacemaker or implantable cardioverter-defibrillator. Med Clin North Am. 2013;97(6):1051-1074.
  10. 10. National Aboriginal Community Controlled Health Organisation (NACCHO). Cardiovascular health in Aboriginal and Torres Strait Islander peoples: position statement. NACCHO, Canberra. 2022.
  11. 11. Boriani G, Fauchier L, Aguinaga L, et al. European Heart Rhythm Association (EHRA) consensus document on management of arrhythmias and cardiac electronic devices in the critically ill and post-surgery patient. Europace. 2019;21(1):7-16.
  12. 12. Australian Commission on Safety and Quality in Health Care (ACSQHC). National Safety and Quality Health Service Standards. 2nd edition. ACSQHC, Sydney. 2017.